Assay
Tetrabutylammonium hydrogen sulfate solution
Dissolve 1 g of tetrabutylammonium hydrogen sulfate in 100 mL of water. Adjust with 1 N sodium hydroxide to a pH of 7.5.
Edetate disodium solution
Dissolve 0.04 g of edetate disodium in 100 mL of water. Adjust with 1 N sodium hydroxide to a pH of 7.5.
pH 7.5 Phosphate buffer
Prepare a mixture of 0.33 M dibasic potassium phosphate and 0.33 M monobasic sodium phosphate (85:15). Adjust, if necessary, by adding more of the appropriate component to a pH of 7.5.
Mobile phase
Transfer, with the aid of 200 mL of water, 50 g of tertiary butyl alcohol to a 1000-mL volumetric flask. Add 60 mL of
pH 7.5 Phosphate buffer, 50 mL of
Tetrabutylammonium hydrogen sulfate solution, and 10 mL of
Edetate disodium solution, and dilute with water to volume. Degas before use. Make adjustments if necessary (see
System Suitability under
Chromatography 621).
Standard preparation
Dissolve an accurately weighed quantity of
USP Oxytetracycline RS in 0.01 N hydrochloric acid to obtain a solution having a known concentration of about 0.22 mg per mL.
System suitability solution
Prepare a solution of tetracycline hydrochloride in 0.01 N hydrochloric acid containing about 0.2 mg per mL. Mix 3 mL of this solution and 1.5 mL of the
Standard preparation, and dilute with water to 25 mL.
Assay preparation
Transfer about 44 mg of Oxytetracycline to a 200-mL volumetric flask, add about 25 mL of 0.01 N hydrochloric acid, swirl to dissolve, dilute with 0.01 N hydrochloric acid to volume, and mix.
Chromatographic system
(see
Chromatography 621)The liquid chromatograph is equipped with a 254-nm detector and a 4.6-mm × 25-cm column that contains packing L21 and is maintained at 60 ± 2
. The flow rate is about 1 mL per minute. Chromatograph the
System suitability solution, and record the peak responses as directed for
Procedure: the relative retention times are about 0.6 for oxytetracycline and 1.0 for tetracycline; and the resolution,
R, between the oxytetracycline peak and the tetracycline peak is not less than 5. Chromatograph the
Standard preparation, and record the peak responses as directed for
Procedure: the tailing factor is not more than 1.25; and the relative standard deviation for replicate injections is not more than 1.0%.
Procedure
[NOTEUse peak areas where peak responses are indicated.
] Separately inject equal volumes (about 20 µL) of the
Standard preparation and the
Assay preparation into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Calculate the quantity, in µg, of C
22H
24N
2O
9 in each mg of the Oxytetracycline taken by the formula:
200(CP / W)(rU / rS),
in which
C is the concentration, in mg per mL, of
USP Oxytetracycline RS in the
Standard preparation; P is the assigned potency, in µg per mg, of
USP Oxytetracycline RS;
W is the weight, in mg, of the Oxytetracycline taken to prepare the
Assay preparation; and
rU and
rS are the oxytetracycline peak responses obtained from the
Assay preparation and the
Standard preparation, respectively.