Procedure
Transfer 4 to 5 mEq of
Internal standard, accurately weighed, to a semimicro sampling tube, add 2 to 3 mL of carbon tetrachloride, apply a sampling valve and septum,
* thereby sealing the tube, and determine the weight of the sealed assembly. Open the valve, introduce about 500 µL of Amyl Nitrite with a syringe, close the valve, and determine the weight of the sealed assembly when it has attained constant weight. Shake the sampling tube and valve assembly, and transfer about 500 µL of the solution to a precision NMR tube as directed for
Absolute Method of Quantitation under
Nuclear Magnetic Resonance 761. With no spinning, or with the spinning adjusted so that the spinning side bands of neither the substance under assay nor the
Internal standard interfere with the regions to be integrated, record as
AS the average area of the
Internal standard singlet appearing at about 5.3 ppm, representing the methylene protons of benzyl benzoate, and record as
AU the average area of the multiplet with a band center at about 4.8 ppm, representing the alpha methylene protons of amyl nitrite, with reference to the tetramethylsilane singlet at 0 ppm. Calculate the quantity of C
5H
11NO
2 in the Amyl Nitrite taken, using 58.57 as the equivalent weight of amyl nitrite (
EWU) and 106.12 as that of benzyl benzoate (
EWS).