Related compounds
Buffer solution
Prepare as directed in the Assay.
Solution A
Prepare a filtered and degassed mixture of Buffer solution and acetonitrile (80:20).
Solution B
Prepare a filtered and degassed mixture of acetonitrile and Buffer solution (60:40).
Mobile phase
Use variable mixtures of
Solution A and
Solution B as directed under
Chromatographic system. Make adjustments if necessary (see
System Suitability under
Chromatography 621).
Diluting solution
Prepare a filtered mixture of 0.02 M sodium acetate, previously adjusted with acetic acid to a pH of 4.0, and acetonitrile (70:30).
Standard solution
Dissolve an accurately weighed quantity of
USP Etoposide RS in
Diluting solution to obtain a Standard stock solution having a known concentration of about 2.0 mg per mL. Dilute this Standard stock solution quantitatively and stepwise with
Diluting solution to obtain a solution having a known concentration of about 10 µg per mL.
System suitability solution
Transfer about 20 mg of n-propylparaben, accurately weighed, to a 100-mL volumetric flask, and dissolve in and dilute with Diluting solution to volume. Transfer 5 mL of this solution and 5 mL of the Standard stock solution to a 50-mL volumetric flask, and dilute with Diluting solution to volume. Transfer 5.0 mL of this solution to a 100-mL volumetric flask, dilute with Diluting solution to volume, and mix.
Test solution
Transfer about 100 mg of Etoposide, accurately weighed, to a 50-mL volumetric flask, and dissolve in and dilute with Diluting solution to volume.
Chromatographic system (see Chromatography 621)
The liquid chromatograph is equipped with a 254-nm detector and a 3.9-mm × 15-cm column that contains packing L11 having a diameter of less than 5 µm. The flow rate is about 1.5 mL per minute. Chromatograph the
System suitability solution using
Solution A, and record the peak responses as directed for
Procedure: the relative retention times are about 0.20 for lignan P, 1.0 for etoposide, and 1.43 for picroetoposide; and the resolution,
R, between propylparaben and etoposide is not less than 1.1. The chromatograph is programmed for
Procedure as follows.
Time (minutes) |
Solution A
(%) |
Solution B
(%) |
Elution |
0 |
100 |
0 |
equilibration |
015 |
100 |
0 |
isocratic |
1530 |
100®40 |
0®60 |
linear gradient |
3040 |
40 |
60 |
isocratic |
4042 |
40®0 |
60®100 |
linear gradient |
4245 |
0 |
100 |
isocratic |
4547 |
0®100 |
100®0 |
linear gradient |
4750 |
100 |
0 |
re-equilibration |
Procedure
Separately inject equal volumes (about 25 µL) of the
Standard solution and the
Test solution into the chromatograph, record the chromatograms for at least 40 minutes, and measure the peak responses. Calculate the percentages of lignan P and picroetoposide in the portion of Etoposide taken by the formula:
5000(C/W)(ri / rS),
in which
C is the concentration, in mg per mL, of
USP Etoposide RS in the
Standard solution; W is the weight, in mg, of Etoposide taken to prepare the
Test solution; ri is the peak response for each related compound obtained from the
Test solution; and
rS is the peak response for etoposide obtained from the
Standard solution: not more than 0.5% of lignan P and 1.0% of picroetoposide is found. Calculate the quantity of any other impurity observed in the chromatogram of the
Test solution by the same formula: not more than 2.0% of all related compounds and other impurities is found.
Assay
Buffer solution
Dissolve 5.44 g of sodium acetate in 2000 mL of water, adjust with glacial acetic acid to a pH of 4.0, and filter.
Mobile phase
Prepare a filtered and degassed mixture of
Buffer solution and acetonitrile (74:26). Make adjustments if necessary (see
System Suitability under
Chromatography 621).
Standard preparation
Dissolve an accurately weighed quantity of
USP Etoposide RS in acetonitrile to obtain a Standard stock solution having a known concentration of about 2.0 mg per mL. Transfer 5.0 mL of this Standard stock solution to a 50-mL volumetric flask, dilute with
Mobile phase to volume, and mix.
System suitability solution
Dissolve an accurately weighed quantity of
USP Etoposide Resolution Mixture RS in
Mobile phase to obtain a solution having a known concentration of 0.3 mg per mL.
Assay preparation
Transfer about 100 mg of Etoposide, accurately weighed, to a 50-mL volumetric flask, dissolve in and dilute with acetonitrile to volume, and mix. Transfer 5.0 mL of this solution to a 50-mL volumetric flask, dilute with Mobile phase to volume, and mix.
Chromatographic system (see Chromatography 621)
The liquid chromatograph is equipped with a 254-nm detector and a 3.9-mm × 30-cm column that contains packing L11. The flow rate is about 1 mL per minute. Chromatograph the
System suitability solution, and record the peak responses as directed for
Procedure: the resolution,
R, between the etoposide and
-etoposide peaks is not less than 1.35. Chromatograph the
Standard preparation, and record the peak responses as directed for
Procedure: the relative standard deviation for replicate injections is not more than 2.0%.
Procedure
Separately inject equal volumes (about 20 µL) of the
Standard preparation and the
Assay preparation into the chromatograph, and allow the
Assay preparation to elute for not less than 1.5 times the retention time of etoposide. Record the chromatograms, and measure the responses for all the peaks. Calculate the quantity, in mg, of C
29H
32O
13 in the portion of Etoposide taken by the formula:
500C(rU / rS),
in which
C is the concentration, in mg per mL, of
USP Etoposide RS in the
Standard preparation; and
rU and
rS are the etoposide peak responses obtained from the
Assay preparation and the
Standard preparation, respectively.