Identification
A:
The retention times of the quinic acid, malic acid, and citric acid peaks in the chromatogram of the
Test preparation correspond to those in the chromatogram of the
Standard preparation, as obtained in the test for
Content of organic acids.
B:
Absence of adulterants
Standard solution
Dissolve accurately weighed quantities of tartaric acid and fumaric acid in water to obtain a solution having concentrations of about 1.0 and 0.1 mg per mL, respectively.
Test solution
Use the Liquid Preparation.
Mobile phase and Chromatographic system
Proceed as directed for Content of organic acids.
Procedure
Separately inject equal volumes (about 20 µL) of the Standard solution and the Test solution into the chromatograph, and record the chromatograms. The retention times of the tartaric acid and fumaric acid peaks in the chromatogram of the Standard solution do not correspond to any of the retention times for peaks observed in the chromatogram of the Test solution.
Limit of sorbitol and sucrose
Mobile phase and Test preparation
Prepare as directed for Content of dextrose and fructose.
Standard preparation
Dissolve accurately weighed quantities of
USP Sorbitol RS and
USP Sucrose RS in water to obtain a solution having known concentrations of about 0.5 mg of each USP Reference Standard per mL.
Chromatographic system
Proceed as directed for Content of dextrose and fructose. Chromatograph the Standard preparation, and record the peak responses as directed for Procedure: the relative retention times are about 0.4 for sucrose and 1.0 for sorbitol; the resolution, R, between sucrose and sorbitol peaks is not less than 1.8; and the relative standard deviation for replicate injections is not more than 2.0%.
Procedure
Separately inject equal volumes (about 20 µL) of the
Standard preparation and the
Test preparation into the chromatograph, record the chromatograms, and measure the areas of the peaks. Calculate the percentages of sucrose and sorbitol in the volume of Liquid Preparation taken by the formula:
0.5(C/V)(rU / rS),
in which
C is the concentration, in mg per mL, of the appropriate USP Reference Standard in the
Standard preparation; V is the volume, in mL, of Liquid Preparation taken for the
Test preparation; and
rU and
rS are the peak responses of the appropriate analyte obtained from the
Test preparation and the
Standard preparation, respectively: not more than 0.05% each of sorbitol and sucrose is found.
Content of dextrose and fructose
Mobile phase
Use filtered and degassed water.
Standard preparation
Dissolve accurately weighed quantities of
USP Dextrose RS and
USP Fructose RS in water to obtain a solution having known concentrations of about 6.0 and 2.0 mg per mL, respectively.
Test preparation
Transfer about 1.0 g of sodium carboxylate cation-exchange resin, accurately weighed, to a 50-mL beaker, add 5 mL of water to make a slurry, and transfer the slurry to a polypropylene automatic pipet fitted with a small plug of silanized glass wool. Quantitatively transfer the slurry to a small chromatographic tube, rinsing the beaker with water and packing the column evenly. Keep the column wet until ready for use. Using a volumetric pipet, transfer 1.0 mL of Liquid Preparation to the column, collect the eluate, and discard it. Pipet 4.0 mL of water onto the top of the column, collect the eluate in a clean vial, and filter if necessary.
Chromatographic system (see Chromatography 621)
The liquid chromatograph is equipped with a refractive index detector and a 7.8-mm × 30-cm analytical column that contains packing L19 and is fitted with a guard column that contains packing L19. The analytical column temperature is maintained at 85
, and the flow rate is about 0.6 mL per minute. Chromatograph the
Standard preparation, and record the peak responses as directed for
Procedure: the relative retention times are about 0.8 for dextrose and 1.0 for fructose; the resolution,
R, between the dextrose and fructose peaks is not less than 1.8; and the relative standard deviation for replicate injections is not more than 2.0%.
Procedure
Separately inject equal volumes (about 20 µL) of the
Standard preparation and the
Test preparation into the chromatograph, record the chromatograms, and measure the areas of all the peaks. Calculate the percentages of dextrose and fructose in the volume of Liquid Preparation taken by the formula:
0.5(C/V)(rU / rS),
in which
C is the concentration, in mg per mL, of the appropriate USP Reference Standard in the
Standard preparation; V is the volume, in mL, of Liquid Preparation taken for the
Test preparation; and
rU and
rS are the peak area responses of the appropriate analyte obtained from the
Test preparation and the
Standard preparation, respectively: not less than 2.4% dextrose and not less than 0.7% of fructose are found.
Content of organic acids
Mobile phase
Transfer about 27.2 g of monobasic potassium phosphate, accurately weighed, to a 1000-mL volumetric flask, and dissolve in 950 mL of water. Adjust with phosphoric acid to a pH of 2.4, dilute with water to volume, mix, and filter.
Test preparation
Use the filtered Liquid Preparation.
Chromatographic system (see Chromatography 621)
The liquid chromatograph is equipped with a 214-nm detector, a 4.6-mm × 25-cm analytical column containing packing L1, and a guard column containing 5-µm packing L1. The flow rate is about 0.6 mL per minute. Prior to use, condition the column with methanol, with water, and finally with
Mobile phase. Chromatograph the
Standard preparation, and record the peak responses as directed for
Procedure: the relative retention times are about 0.4 for quinic acid, 0.5 for malic acid, and 1.0 for citric acid; the resolution,
R, between quinic acid and malic acid is not less than 2.5; and the relative standard deviation for replicate injections is not more than 2.0%.
Procedure
Separately inject equal volumes (about 20 µL) of the
Standard preparation and the
Test preparation into the chromatograph, record the chromatograms, and measure the peak area responses. Calculate the percentages of quinic acid, malic acid, and citric acid in the Liquid Preparation by the formula:
0.1C(rU / rS),
in which
C is the concentration, in mg per mL, of the appropriate USP Reference Standard in the
Standard preparation; and
rU and
rS are the peak area responses of the appropriate analyte obtained from the
Test preparation and the
Standard preparation, respectively: not less than 0.9% each of quinic acid and citric acid is found; not less than 0.7% of malic acid is found; and the ratio of quinic acid to malic acid is not less than 1.0.