Related compounds
Phosphoric acid solution
Transfer 6.0 mL of phosphoric acid to a 50-mL volumetric flask, and dilute with water to volume.
Buffer solution
Dissolve about 3.5 g of sodium 1-pentanesulfonate in 1000 mL of water, and adjust with Phosphoric acid solution or 1 N sodium hydroxide to a pH of 3.0 ± 0.05.
Mobile phase
Prepare a filtered and degassed mixture of
Buffer solution and acetonitrile (80:20). Make adjustments if necessary (see
System Suitability under
Chromatography 621).
Tizanidine related compound A solution
Dissolve an accurately weighed quantity of USP Tizanidine Related Compound A RS in methanol, and dilute quantitatively, and stepwise if necessary, with methanol to obtain a solution having a known concentration of about 0.1 mg per mL.
Tizanidine related compound B solution
Dissolve an accurately weighed quantity of USP Tizanidine Related Compound B RS in methanol, and dilute quantitatively, and stepwise if necessary, with methanol to obtain a solution having a known concentration of about 0.1 mg per mL.
Tizanidine related compound C solution
Dissolve an accurately weighed quantity of USP Tizanidine Related Compound C RS in methanol, and dilute quantitatively, and stepwise if necessary, with methanol to obtain a solution having a known concentration of about 0.1 mg per mL.
Resolution solution
Transfer about 23 mg of USP Tizanidine Hydrochloride RS to a 100-mL volumetric flask, add 20 mL of Mobile phase and 10 mL each of Tizanidine related compound A solution, Tizanidine related compound B solution, and Tizanidine related compound C solution. Sonicate to dissolve the USP Tizanidine Hydrochloride RS, and dilute with Mobile phase to volume.
Standard solution
Dissolve an accurately weighed quantity of USP Tizanidine Hydrochloride RS in Mobile phase, and dilute quantitatively, and stepwise if necessary, with Mobile phase to obtain a solution having a known concentration of about 0.046 mg per mL.
Test solution
Transfer about 57 mg of Tizanidine Hydrochloride, accurately weighed, to a 50-mL volumetric flask, dissolve in and dilute with Mobile phase to volume, and mix.
Chromatographic system (see Chromatography 621)
The liquid chromatograph is equipped with a 230-nm detector and a 4.6-mm × 25-cm column that contains packing L1. The flow rate is about 1.0 mL per minute. The column temperature is maintained at 50
. Chromatograph the
Resolution solution, and record the peak responses as directed for
Procedure: the relative retention times are given in
Table 1; the resolution,
R, between tizanidine and tizanidine related compound C is not less than 4.0; and the resolution,
R, between tizanidine and tizanidine related compound B is not less than 4.0. Chromatograph the
Standard solution, and record the peak responses as directed for
Procedure: the column efficiency is not less than 5000 theoretical plates; the tailing factor is not more than 2.0; and the relative standard deviation for replicate injections is not more than 2.0%.
Procedure
Inject equal volumes (about 10 µL) of the
Standard solution and the
Test solution into the chromatograph, record the chromatograms, and measure the responses for the major analyte peaks, disregarding the peaks due to the solvent. Calculate the percentage of each impurity in the portion of Tizanidine Hydrochloride taken by the formula:
(253.71/290.17)5000(C/W)F(ri / rS)
in which 253.71 and 290.17 are the molecular weights of tizanidine and tizanidine hydrochloride, respectively;
C is the concentration, in mg per mL, of USP Tizanidine Hydrochloride RS in the
Standard solution; W is the weight, in mg, of sample taken to prepare the
Test solution; F is the relative response factor and is given in
Table 1; ri is the peak area for each impurity obtained from the
Test solution; and
rS is the peak area of tizanidine obtained from the
Standard solution. The limits for the impurities are specified in
Table 1.
Table 1
Compound Name |
Relative Retention Time |
Relative Response Factor |
Limit (%) |
Tizanidine related com- pound C |
about 0.8 |
1.0 |
0.1 |
Tizanidine |
1.0 |
|
|
Tizanidine related com- pound B |
about 1.4 |
0.9 |
0.1 |
Tizanidine related com- pound A |
about 10.2 |
0.9 |
0.1 |
Individual unknown |
|
1.0 |
0.1 |
Total |
|
|
0.3 |
Assay
Buffer solution
Dissolve 6.8 g of monobasic potassium phosphate in 1000 mL of water, and adjust with 5.3 N potassium hydroxide to a pH of 7.5 ± 0.05.
Mobile phase
Prepare a filtered and degassed mixture of
Buffer solution and acetonitrile (1:1). Make adjustments if necessary (see
System Suitability under
Chromatography 621).
System suitability preparation
Dissolve suitable quantities of USP Tizanidine Hydrochloride RS and USP Tizanidine Related Compound C RS in Mobile phase, and dilute quantitatively, and stepwise if necessary, with Mobile phase to obtain a solution containing about 0.00012 mg per mL and 0.046 mg per mL, respectively.
Standard preparation
Dissolve an accurately weighed quantity of USP Tizanidine Hydrochloride RS in Mobile phase, and dilute quantitatively, and stepwise if necessary, with Mobile phase to obtain a solution having a known concentration of about 0.046 mg per mL.
Assay preparation
Transfer about 23 mg of Tizanidine Hydrochloride, accurately weighed, to a 100-mL volumetric flask, dissolve in and dilute with Mobile phase to volume, and mix. Transfer 10.0 mL of this solution to a 50-mL volumetric flask, dilute with Mobile phase to volume, and mix.
Chromatographic system (see Chromatography 621)
The liquid chromatograph is equipped with a 230-nm detector and a 4.6-mm × 15-cm column that contains packing L7. The flow rate is about 1.0 mL per minute. The column temperature is maintained at 35
. Chromatograph the
System suitability preparation, and record the peak responses as directed for
Procedure: the relative retention times are about 0.5 for tizanidine related compound C and 1.0 for tizanidine; the resolution,
R, between tizanidine and tizanidine related compound C is not less than 13.0; and the tailing factor for the tizanidine peak is not more than 1.6. Chromatograph the
Standard preparation, and record the peak responses as directed for
Procedure: the relative standard deviation for replicate injections is not more than 2.0%.
Procedure
Separately inject equal volumes (about 20 µL) of the
Standard preparation and the
Assay preparation into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Calculate the quantity, in mg, of C
9H
8ClN
5S·HCl in the portion of Tizanidine Hydrochloride taken by the formula:
500C(rU / rS)
in which
C is the concentration, in mg per mL, of USP Tizanidine Hydrochloride RS in the
Standard preparation; and
rU and
rS are the peak areas obtained from the
Assay preparation and the
Standard preparation, respectively.