Packaging and storage
Preserve in well-closed containers. No storage requirements specified.
Labeling
The labeling indicates the percentage content, on the anhydrous basis, of D-sorbitol and 1,4-sorbitan.
Identification
A:
Prepare a solution containing 1.4 g of Sorbitol Sorbitan Solution in 75 mL of water. Transfer 3 mL of this solution to a 15-cm test tube, add 3 mL of freshly prepared catechol solution (1 in 10), and mix. Add 6 mL of sulfuric acid, mix again, then gently heat the tube in a flame for about 30 seconds: a deep pink or wine-red color appears.
B:
The retention times of the major peaks in the chromatogram of the Assay preparation correspond to those in the chromatogram of the Standard preparation, as obtained in the Assay.
Microbial limits 61
The total aerobic microbial count using the
Plate Method is not more than 1000 cfu per mL. The total combined molds and yeasts count is not more than 100 cfu per mL.
pH 791:
between 4.0 and 7.0, in a 14% (w/w) solution of Sorbitol Sorbitan Solution in carbon dioxide-free water.
Residue on ignition 281:
not more than 0.20%, calculated on the anhydrous basis. Determine on a 2-g portion, accurately weighed.
Reducing sugars
To an amount of Sorbitol Sorbitan Solution, equivalent to 3.3 g, on the anhydrous basis, add 3 mL of water, 20.0 mL of cupric citrate TS, and a few glass beads. Heat so that boiling begins after 4 minutes, and maintain boiling for 3 minutes. Cool rapidly, and add 40 mL of diluted acetic acid, 60 mL of water, and 20.0 mL of 0.05 N iodine VS. With continuous shaking, add 25 mL of a mixture of 6 mL of hydrochloric acid and 94 mL of water. When the precipitate has dissolved, titrate the excess of iodine with 0.05 N sodium thiosulfate VS using 2 mL of starch TS, added towards the end of the titration, as an indicator. Not less than 12.8 mL of 0.05 N sodium thiosulfate VS is required, corresponding to not more than 0.3% of reducing sugars, on the anhydrous basis, as glucose. The amount determined in this test is not included in the calculated amount under Other Impurities.
Limit of nickel
Test solution
Dissolve 20.0 g of Sorbitol Sorbitan Solution in diluted acetic acid, and dilute with diluted acetic acid to 100.0 mL. Add 2.0 mL of a saturated solution of ammonium pyrrolidine dithiocarbamate (about 10 g of ammonium pyrrolidine dithiocarbamate per L) and 10.0 mL of methyl isobutyl ketone, and shake for 30 seconds. Protect from bright light. Allow the two layers to separate, and use the methyl isobutyl ketone layer.
Blank solution
Prepare as directed for the Test solution, except to omit the use of Sorbitol Sorbitan Solution. Quantities should be increased five fold to ensure that a sufficient volume of Blank solution is available.
Standard solutions
Prepare as directed for the Test solution, except to prepare three solutions by adding 0.5 mL, 1.0 mL, and 1.5 mL of nickel standard solution TS.
Procedure
Set the instrument to zero using the
Blank solution. Concomitantly determine the absorbances of the
Standard solutions and the
Test solution at least three times each, at the wavelength of maximum absorbance at 232.0 nm, with a suitable atomic absorption spectrophotometer (see
Spectrophotometry and Light-Scattering 851) equipped with a nickel hollow-cathode lamp and an airacetylene flame. Record the average of the steady readings for each of the
Standard solutions and the
Test solution. Between each measurement, aspirate the
Blank solution, and ascertain that the reading returns to zero. Plot the absorbances of the
Standard solutions and the
Test solution versus the added quantity of nickel. Extrapolate the line joining the points on the graph until it meets the concentration axis. The distance between this point and the intersection of the axes represents the concentration of nickel in the
Test solution. Not more than 1 µg per g, calculated on the anhydrous basis, is found.
Residual solvents 467:
meets the requirements.
(Official January 1, 2007)
Assay
Mobile phase
Use degassed water.
Resolution solution
Dissolve sorbitol, 1,4-sorbitan, isosorbide, and mannitol in water to obtain a solution having concentrations of about 10 mg per g, 4 mg per g, 4 mg per g, and 1 mg per g, respectively.
Standard preparation
Dissolve accurately weighed quantities of
USP Sorbitol RS and
USP 1,4-Sorbitan RS in water to obtain a solution having concentrations of about 10 mg per g and 4 mg per g, respectively.
Assay preparation
Dissolve about 0.40 g of Sorbitol Sorbitan Solution, accurately weighed, in water, and dilute with water to about 20 g. Accurately record the final solution weight, and mix thoroughly.
Chromatographic system (see Chromatography 621)
The liquid chromatograph is equipped with a refractive index detector that is maintained at a constant temperature of about 35
, and a 7.8-mm × 10-cm column that contains packing L34. The column temperature is maintained at about 50
, controlled within ±2
, and the flow rate is about 0.6 mL per minute. Chromatograph the
Resolution solution, and record the peak responses as directed for
Procedure: the resolution,
R, between the 1,4-sorbitan and isosorbide is not less than 2.0. Chromatograph the
Standard preparation, and record the peak responses for 1,4-sorbitan and sorbitol as directed for
Procedure: the relative retention times are about 0.35 for 1,4-sorbitan, 0.43 for isosorbide, 0.7 for mannitol, and 1.0 for sorbitol; and the relative standard deviation for replicate injections is not more than 2.0% for each analyte.
Procedure
Separately inject equal volumes (about 10 µL) of the
Standard preparation and the
Assay preparation into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Separately calculate the percentages, on the anhydrous basis, of 1,4-sorbitan and
D-sorbitol in the portion of Sorbitol Sorbitan Solution taken by the formula:
[10,000(CS / CU)(rU / rS)]/(100 W),
in which
CS is the concentration, in mg per g, of the appropriate USP Reference Standard in the
Standard preparation; CU is the concentration, in mg per g, of the Sorbitol Sorbitan Solution in the
Assay preparation; rU and
rS are the peak responses of the corresponding analyte obtained from the
Assay preparation and the
Standard preparation, respectively; and
W is the percentage obtained in the test for
Water.