Procedure
Separately apply 5 µL of the
Test solution and 5 µL of each of the
Standard solutions to a thin-layer chromatographic plate (see
Chromatography 621) coated with a 0.25-mm layer of chromatographic silica gel. Allow the spots to dry, and develop the chromatogram in a solvent system consisting of a mixture of cyclohexane, ethyl acetate, and ether (170:40:25) in a paper-lined tank until the solvent front has moved about three-fourths of the length of the plate. Remove the plate from the developing chamber, mark the solvent front, allow to air-dry, and examine the plate under short-wavelength UV light. Compare the intensity of the secondary spot in the chromatogram of the
Test solution having the same
RF value as that of the primary spot of
Standard solution C: the spot is not larger or more intense than the principal spot in the chromatogram of
Standard solution C. Compare the intensities of any additional secondary spots observed in the chromatogram of the
Test solution with that of the principal spot in the chromatogram of
Standard solution B: the sum of the intensities of the additional secondary spots obtained from the
Test solution corresponds to not more than 0.2%.