Chromatographic purity
Mobile phase
Prepare as directed in the
Assay.
Standard solution
Dissolve an accurately weighed quantity of
USP Phenytoin RS in methanol, and dilute quantitatively, and stepwise if necessary, with methanol to obtain a solution having a known concentration of about 10 µg per mL.
Test solution
Use
Assay preparation A, prepared as directed in the
Assay.
Resolution solution
Prepare a solution of benzoin in methanol having a concentration of about 10 µg per mL. Dissolve 10 mg of
USP Phenytoin RS in 10.0 mL of the benzoin solution.
Chromatographic system
(see
Chromatography 621)The liquid chromatograph is equipped with a 220-nm detector and a 4.6-mm × 25-cm column that contains 5-µm packing L1. The flow rate is about 1.5 mL per minute. Chromatograph the
Resolution solution, and record the peak responses as directed for the
Procedure: the relative retention times are about 0.75 for phenytoin and 1.0 for benzoin; and the resolution,
R, is not less than 1.5.
Procedure
[NOTEUse peak areas where peak responses are indicated.
] Separately inject equal volumes (about 20 µL) of the
Standard solution and the
Test solution into the chromatograph, record the chromatograms, and measure the responses for all of the peaks. Calculate the percentage of each impurity peak in the
Test solution taken by the formula:
100(C / D)(ri / rS),
in which
C is the concentration, in µg per mL, of
USP Phenytoin RS in the
Standard solution; D is the concentration, in µg per mL, of phenytoin in the
Test solution; ri is the peak response for each impurity; and
rS is the peak response for phenytoin in the
Standard solution: not more than 0.9% total impurities is found, excluding benzophenone.
Limit of benzophenone
Standard solution
Dissolve an accurately weighed quantity of benzophenone in methanol, and dilute quantitatively, and stepwise if necessary, with methanol to obtain a solution having a known concentration of about 1.0 µg per mL.
Chromatographic system (see Chromatography 621)
Use the same system as directed in the test for
Chromatographic purity except to chromatograph three injections of the
Standard solution, and record the peak responses as directed for
Procedure: the relative standard deviation is not more than 5.0%.
Procedure
[NOTEUse peak areas where peak responses are indicated.
] Separately inject equal volumes (about 20 µL) of the
Standard solution and the
Test solution into the chromatograph, record the chromatograms, and measure the responses for all of the peaks. The relative retention times are about 0.25 for phenytoin and 1.0 for benzophenone. Calculate the percentage of benzophenone in the portion of Phenytoin taken by the formula:
100(C / D)(rU / rS),
in which
C is the concentration, in µg per mL, of benzophenone in the
Standard solution; D is the concentration, in µg per mL, of phenytoin in the
Test solution; and
rU and
rS are the benzophenone peak responses obtained from the
Test solution and the
Standard solution, respectively: not more than 0.1% of benzophenone is found.
Assay
Mobile phase
Prepare a filtered and degassed mixture of methanol and water (55:45). Make adjustments if necessary (see
System Suitability under
Chromatography 621).
Standard preparation
Dissolve, with the aid of sonication if necessary, an accurately weighed quantity of
USP Phenytoin RS in
Mobile phase, and dilute quantitatively, and stepwise if necessary, with
Mobile phase to obtain a solution having a known concentration of about 100 µg per mL.
Assay preparation
Transfer about 100 mg of Phenytoin, accurately weighed, to a 100-mL volumetric flask, dissolve in and dilute with methanol to volume, and mix (Assay preparation A). Transfer 10.0 mL of this solution to a second 100-mL volumetric flask, dilute with Mobile phase to volume, and mix (Assay preparation B).
Resolution solution
Prepare a solution of benzoin in Mobile phase having a concentration of about 1.5 mg per mL. Mix 1.0 mL of this solution and 9.0 mL of the Standard preparation.
Chromatographic system
(see
Chromatography 621)The liquid chromatograph is equipped with a 220-nm detector and a 4.6-mm × 25-cm column that contains 5-µm packing L1. The flow rate is about 1.5 mL per minute. Chromatograph the
Standard preparation, and record the peak responses as directed for
Procedure: the relative standard deviation is not more than 1.0%. Chromatograph the
Resolution solution. The relative retention times are about 0.75 for phenytoin and 1.0 for benzoin; the resolution,
R, is not less than 1.5; and the tailing factor for the phenytoin peak is not more than 1.5.
Procedure
Separately inject equal volumes (about 20 µL) of the
Standard preparation and
Assay preparation B into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Calculate the quantity, in mg, of C
15H
12N
2O
2 in the portion of Phenytoin taken by the formula:
(1000C)(rU / rS),
in which
C is the concentration, in mg per mL, of
USP Phenytoin RS in the
Standard preparation; and
rU and
rS are the peak responses obtained from the
Assay preparation and the
Standard preparation, respectively.