Dissolution 711
Medium:
water; 900 mL.
Apparatus 1:
100 rpm.
Time:
120 minutes.
Procedure
Determine the amount of C
11H
11NO
2 dissolved, employing the procedure set forth in the
Assay, making any necessary modifications.
Tolerances
Not less than 75% (Q) of the labeled amount of C11H11NO2 is dissolved in 120 minutes.
Assay
Mobile phase
Prepare a filtered and degassed mixture of water and acetonitrile (55:45). Make adjustments if necessary (see
System Suitability under
Chromatography 621).
Standard preparation
Dissolve an accurately weighed quantity of
USP Phensuximide RS in
Mobile phase to obtain a solution having a known concentration of about 1 mg per mL.
Assay preparation
Place 10 Capsules in a 500-mL volumetric flask, and add 280 mL of water. Sonicate in a water bath at 40
to 50
, with occasional shaking, until the Capsules have broken, and cool to room temperature. Dilute with acetonitrile to volume, mix, and filter. Transfer an accurately measured volume of this specimen solution, equivalent to about 50 mg of Phensuximide, to a 50-mL volumetric flask, dilute with
Mobile phase to volume, and mix.
Chromatographic system
(see
Chromatography 621)The liquid chromatograph is equipped with a 254-nm detector and a 3.9-mm × 30-cm column that contains packing L1. The flow rate is about 1 mL per minute. Chromatograph the
Standard preparation, and record the peak responses as directed for
Procedure: the column efficiency determined from the analyte peak is not less than 2100 theoretical plates, and the relative standard deviation for replicate injections is not more than 1.5%.
Procedure
Separately inject equal volumes (about 20 µL) of the
Standard preparation and the
Assay preparation into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Calculate the quantity, in mg, of C
11H
11NO
2 per Capsule taken by the formula:
2500(C / V)(rU / rS),
in which
C is the concentration, in mg per mL, of
USP Phensuximide RS in the
Standard preparation,
V is the volume, in mL, of specimen solution taken for the
Assay preparation, and
rU and
rS are the phensuximide peak responses obtained from the
Assay preparation and the
Standard preparation, respectively.