Related compounds
Solution A, Solution B, Mobile phase, and Diluent
Prepare as directed in the Assay.
Test solution
Transfer about 25 mg of Letrozole, accurately weighed, to a 250-mL volumetric flask, dissolve in 75 mL of acetonitrile, dilute with water to volume, and mix.
Reference solution
Transfer 5.0 mL of the Test solution to a 50-mL volumetric flask, and dilute with Diluent to volume. Transfer 5.0 mL of the solution so obtained to a 50-mL volumetric flask, dilute with Diluent to volume, and mix.
Chromatographic system
Prepare as directed in the Assay. Chromatograph the System suitability solution, and record the peak responses as directed for Procedure: the relative retention times are about 0.68 for letrozole related compound A and 1.0 for letrozole; and the resolution, R, between letrozole related compound A and letrozole is not less than 1.5. Chromatograph the Reference solution, and record the peak responses as directed for Procedure: the relative standard deviation for replicate injections is not more than 10.0%.
Procedure
[NOTE4,4
¢,4
¢¢-Methylidenetrisbenzonitrile may be present; its relative retention time is about 1.9.
] Separately inject equal volumes (about 15 µL) of the
Test solution and the
Reference solution into the chromatograph, record the chromatograms, and measure the areas for the major peaks. Calculate the percentage of each impurity in the portion of Letrozole taken by the formula:
ri / rR,
in which
ri is the peak response of letrozole related compound A, 4,4
¢,4
¢¢-methylidenetrisbenzonitrile, or any other impurity obtained from the
Test solution; and
rR is the peak response of letrozole obtained from the
Reference solution: not more than 0.3% of letrozole related compound A is found; not more than 0.2% of 4,4
¢,4
¢¢-methylidenetrisbenzonitrile is found; not more than 0.1% of any other impurity is found; and not more than 0.3% of all other impurities is found.
Assay
Solution A
Use filtered and degassed water.
Solution B
Use filtered and degassed acetonitrile.
Mobile phase
Use variable mixtures of
Solution A and
Solution B as directed for
Chromatographic system. Make adjustments if necessary (see
System Suitability under
Chromatography 621).
Diluent
Prepare a filtered and degassed mixture of Solution A and Solution B (7:3).
Standard preparation
Dissolve an accurately weighed quantity of
USP Letrozole RS in
Diluent, and dilute quantitatively, and stepwise if necessary, with
Diluent to obtain a solution having a known concentration of about 10 µg per mL.
Assay preparation
Transfer about 20 mg of Letrozole, accurately weighed, to a 100-mL volumetric flask. Dissolve in and dilute with Diluent to volume, and mix. Pipet 5.0 mL of this solution into a 100-mL volumetric flask, dilute with Diluent to volume, and mix.
Chromatographic system (see Chromatography 621)
The liquid chromatograph is equipped with a 230-nm detector and a 4.6-mm × 12.5-cm column that contains 5-µm packing L1. The flow rate is about 1.0 mL per minute. The chromatograph is programmed as follows.
Time (minutes) |
Solution A
(%) |
Solution B
(%) |
Elution |
0 |
70 |
30 |
equilibration |
025 |
70®30 |
30®70 |
linear gradient |
Chromatograph the
Standard preparation, and record the peak responses as directed for
Procedure: the tailing factor is between 0.8 and 1.5; and the relative standard deviation for replicate injections is not more than 2.0%.
Procedure
Separately inject equal volumes (about 20 µL) of the
Standard preparation and the
Assay preparation into the chromatograph, record the chromatograms, and measure the areas for the major peaks. Calculate the quantity, in mg, of C
17H
11N
5 in the portion of Letrozole taken by the formula:
2000C(rU / rS),
in which
C is the concentration, in mg per mL, of
USP Letrozole RS in the
Standard preparation; and
rU and
rS are the peak responses obtained from the
Assay preparation and the
Standard preparation, respectively.