Identification
A: 
Pipet 1 mL of Injection into a 125-mL separator containing 10 mL of water and 1 mL of 1 N sodium hydroxide, mix, extract with two 10-mL portions of methylene chloride, and evaporate the extracts on a steam bath just to dryness. Dissolve the residue in methanol, add 1 mL of 1.2 N hydrochloric acid, then add methanol to make 100 mL. Dilute 10 mL of this solution with methanol to 100 mL: the UV absorption spectrum of this solution exhibits a maximum at the same wavelength as that of a similar solution of 
USP Amitriptyline Hydrochloride RS, concomitantly measured.
 
B: 
The retention time of the major peak in the chromatogram of the 
Assay preparation corresponds to that in the chromatogram of the 
Standard preparation, as obtained in the 
Assay.
 
C: 
Pipet a volume of Injection, equivalent to 50 mg of amitriptyline hydrochloride, into a separator containing 25 mL of water. Proceed as directed in the test for 
IdentificationOrganic Nitrogenous Bases 
181
, beginning with In a second separator, and using water in place of 0.01 N hydrochloric acid in the Reference Standard solution: the solution of the test specimen so obtained meets the requirements of the test for 
IdentificationOrganic Nitrogenous Bases 
181
.
 
 
Pyrogen 
151
Amitriptyline Hydrochloride Injection, diluted with Sodium Chloride Injection containing 0.9 percent of NaCl to a concentration of 2.5 mg of amitriptyline hydrochloride per mL, meets the requirements, the test dose being 1 mL per kg.
 
Assay
Phosphate buffer
Dissolve 11.04 g of monobasic sodium phosphate in 900 mL of water, adjust with phosphoric acid to a pH of 2.5 ± 0.5, dilute with water to make 1000 mL, and mix.
Mobile phase
Prepare a filtered and degassed mixture of 
Phosphate buffer and acetonitrile (58:42). Make adjustments if necessary (see 
System Suitability under 
Chromatography 
621
).
 
Standard preparation
Dissolve an accurately weighed quantity of 
USP Amitriptyline Hydrochloride RS in water, and dilute quantitatively with water to obtain a solution having a known concentration of about 0.2 mg per mL.
 
Assay preparation
Transfer an accurately measured volume of Injection, equivalent to about 50 mg of amitriptyline hydrochloride, to a 250-mL volumetric flask, dilute with water to volume, and mix.
Chromatographic system
 (see 
Chromatography 
621
)The liquid chromatograph is equipped with a 254-nm detector and a 4-mm × 30-cm column that contains packing L1. The flow rate is about 2 mL per minute. Chromatograph the 
Standard preparation, and record the peak responses as directed for 
Procedure: the tailing factor is not more than 2.0, the column efficiency is not less than 800 theoretical plates, and the relative standard deviation for replicate injections is not more than 2.0%.
 
Procedure
Separately inject equal volumes (about 20 µL) of the 
Standard preparation and the 
Assay preparation into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Calculate the quantity, in mg, of amitriptyline hydrochloride (C
20H
23N·HCl) in each mL of the Injection taken by the formula: 
250(C / V)(rU / rS),
in which 
C is the concentration, in mg per mL, of 
USP Amitriptyline Hydrochloride RS in the 
Standard preparation; V is the volume, in mL, of Injection taken; and 
rU and 
rS are the peak responses obtained from the 
Assay preparation and the 
Standard preparation, respectively.