Specific activity:
no carrier added.
Packaging and storage
Preserve in single-dose or multiple-dose containers that are adequately shielded.
Labeling
Label container to include the following, in addition to the information specified for
Labeling under
Injections 1: the time and date of calibration; the amount of
18F as fluoride anion expressed as MBq (or mCi) per mL, at time of calibration; total activity at time of calibration; the expiration time and date; and the statements CautionRadioactive Material and Do not use if cloudy or if it contains particulate matter. Label it to indicate that the radioactive half-life of
18F is 109.7 minutes, and that in making dosage calculations, correction is to be made for radioactive decay.
Radionuclide identification (see Radioactivity 821)
A:
Its half-life, determined using a suitable detector system, is between 105 and 115 minutes.
B:
Radiochemical identityThe retention time of the major peak in the chromatogram of the Test solution corresponds to that in the chromatogram of the Standard solution, as obtained in the test for Radiochemical purity.
Chemical purity
This article may be synthesized by different methods and processes and may, therefore, may contain different impurities. The presence of unlabeled ingredients, reagents, and by-products specific to the process must be controlled, and their potential for physiological or pharmacological effects must be considered.
Radiochemical purity
Mobile phase
Prepare a filtered and degassed solution of 0.003 N sulfuric acid in water.
Standard solution
Dissolve an accurately weighed quantity of
USP Sodium Fluoride RS in water, and dilute quantitatively, and stepwise if necessary, with water to obtain a solution having a known concentration of about 0.1 mg per mL.
Test solution
Use the Injection.
Chromatographic system
(see
Chromatography 621)The liquid chromatograph is equipped with a flow-through gamma ray detector, a conductivity detector, and a 7.5-mm × 30-cm column that contains 10-µm packing L17. The flow rate is about 0.8 mL per minute. Chromatograph the
Test solution, and record the peak responses as directed for
Procedure: the relative standard deviation for replicate injections is not more than 5%.
Procedure
Prepare a solution of the Test solution and the Standard solution, inject about 50 µL of the combined solution into the chromatograph, record the chromatograms, and measure the areas for both the radioactive and nonradioactive peaks (the injection volume may be adjusted to obtain suitable detection system sensitivity): the radioactivity of the major peak is not less than 95% of the total radioactivity measured; and the retention time of the Test solution corresponds to the retention time (about 8 minutes) of the Standard solution.
Radionuclide purity
Using a suitable gamma-ray spectrometer (see
Selection of a Counting Assembly under
Radioactivity 821), count an appropriate aliquot of Injection for a period of time sufficient to obtain a gamma spectrum. The resultant gamma spectrum should be analyzed for the presence of identifiable photopeaks which are not characteristic of
18F emission: not less than 99.5% of the gamma emissions correspond to the 0.511 MeV, 1.022 MeV, or Compton scatter peaks of
18F.
Other requirements
It meets the requirements under
Injections 1, except that the Injection may be distributed or dispensed prior to completion of the test for
Sterility 71, the latter test started within 24 hours of final manufacture, and except that it is not subject to the recommendation on
Volume in Container.
Assay for radioactivity
Using a suitable calibrated system as directed under
Radioactivity 821, determine the radioactivity, in MBq (or MCi) per mL, of Injection.