Identification
B:
Prepare a test solution in methanol containing 1 mg per mL. Apply 1 µL of the test solution and 1 µL of a Standard solution of
USP Digitoxin RS in methanol containing 1 mg per mL to a suitable thin-layer chromatographic plate (see
Chromatography 621) coated with a 0.25-mm layer of chromatographic silica gel mixture. Allow the applications to dry, and develop the chromatogram in a solvent system consisting of a mixture of methylene chloride and methanol (93:7) until the solvent front has moved about three-fourths of the length of the plate. Remove the plate from the developing chamber, mark the solvent front, and dry the plate at 100
to remove the solvent. Spray the plate with a 6 in 10 solution of sulfuric acid in methanol, heat at 105
for 10 minutes, and examine the chromatogram under long-wavelength UV light: the
RF value of the principal spot obtained from the test solution corresponds to that obtained from the Standard solution.
C:
The retention time of the major peak in the chromatogram of the
Assay preparation corresponds to that of the major peak in the chromatogram of the
Standard preparation, as obtained in the
Assay.
Assay
Mobile phase
Prepare a filtered and degassed mixture of water and acetonitrile (55:45). Make adjustments if necessary (see
System Suitability under
Chromatography 621).
Standard preparation
Dissolve an accurately weighed quantity of
USP Digitoxin RS in
Mobile phase, and dilute quantitatively, and stepwise if necessary, with
Mobile phase to obtain a solution having a known concentration of about 40 µg per mL.
Assay preparation
Transfer about 50 mg of Digitoxin, accurately weighed, to a 200-mL volumetric flask. Dissolve in and dilute with Mobile phase to volume, and mix. Pipet 4 mL of this solution into a 25-mL volumetric flask, dilute with Mobile phase to volume, and mix.
System suitability preparation
Prepare a solution in Mobile phase containing about 40 µg each of digitoxin and digoxin per mL.
Chromatographic system
(see
Chromatography 621)The liquid chromatograph is equipped with a 218-nm detector and a 3.9-mm × 30-cm column that contains packing L1. The flow rate is about 1 mL per minute. Chromatograph the
Standard preparation and the
System suitability preparation, and record the peak responses as directed for
Procedure: the relative retention times are about 0.35 for digoxin and 1.0 for digitoxin; the resolution,
R, between the digoxin and digitoxin peaks is not less than 2.0; the tailing factor for the analyte peak is not more than 2.0; and the relative standard deviation for replicate injections of the
Standard preparation is not more than 2.0%.
Procedure
Separately inject equal volumes (about 50 µL) of the
Standard preparation and the
Assay preparation into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Calculate the quantity, in mg, of C
41H
64O
13 in the portion of Digitoxin taken by the formula:
1.25C(rU / rS),
in which
C is the concentration, in µg per mL, of
USP Digitoxin RS in the
Standard preparation, and
rU and
rS are the peak responses obtained from the
Assay preparation and the
Standard preparation, respectively.