Drug release 724
Medium:
0.001 M phosphoric acid; 80 mL for systems containing 5 mg or less of clonidine; 200 mL for systems containing more than 5 mg of clonidine.
Times:
8, 24, 96, and 168 hours.
Apparatus 7
Proceed as directed in the chapter, using the transdermal system holder-angled disk (see
Figure 4a). The appropriate size of the holder, 1.42 or 1.98 inches, should be chosen based on the size of the system to prevent overhang. Use 100-mL beakers for
Medium volumes of 80-mL and 300-mL beakers for
Medium volumes of 200 mL. Gently press the transdermal system to a dry, smooth, square piece of cellulose membrane
* , or equivalent, with the adhesive side against the membrane. Attach the membrane/system to a suitable inert sample holder with a Viton O-ring, or equivalent, such that the backing of the system is adjacent to, and centered on, the bottom of the sample holder. Trim the excess of cellulose membrane with scissors. Suspend each sample holder from the arm of a reciprocating shaker such that each system is continuously immersed in a beaker containing the specified volume of
Medium. The filled beakers are weighed and pre-equilibrated to 32.0 ± 0.3
prior to immersing the test sample. Agitate the sample in an up-down motion at a frequency of 30 cycles per minute with an amplitude of 2.0 ± 0.1 cm. The
Medium must be added daily to the beakers during each interval to maintain sample immersion. At the end of each time interval, transfer the test sample to a fresh beaker containing the appropriate volume of
Medium, weighed and pre-equilibrated to 32.0 ± 0.3
.
Determine the amount of C9H9Cl2N3 released by employing the following method.
Mobile phase
Use a filtered and degassed 0.1% solution of triethylamine in a mixture of water and methanol (70:30), adjust with phosphoric acid to a pH of 6.0 ± 0.2. Make adjustments if necessary (see
System Suitability under
Chromatography 621).
System suitability solution
Prepare a solution of USP Clonidine RS in 0.001 M phosphoric acid having a known concentration of about 10 µg per mL.
Standard solutions
Prepare a minimum of four standard solutions of USP Clonidine RS in 0.001 M phosphoric acid having known concentrations of clonidine similar to those of the Test solutions.
Test solutions
At the end of each release interval, allow the beakers to cool to room temperature and make up for evaporative Medium losses by adding Medium to obtain the original weight. Mix.
Chromatographic system (see Chromatography 621)
The liquid chromatograph is equipped with a 220-nm detector and a 4.6-mm × 15-cm column that contains packing L1. The flow rate is about 1.5 mL per minute. Chromatograph the
System suitability solution, and record the peak responses as directed for
Procedure: the tailing factor is not more than 2.0; the capacity factor is not less than 0.5; the column efficiency is not less than 2000 theoretical plates; and the relative standard deviation for replicate injections is not more than 2.0%.
Procedure
Separately inject equal volumes (about 25 µL) of filtered portions of the
Standard solution and the
Test solution into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Construct a standard curve of concentration (µg per mL) of clonidine in the
Standard solutions versus peak area by linear regression analysis. The correlation coefficient is not less than 0.995. Calculate the release rate of clonidine by the formula:
CV/TA,
in which
C is the concentration, in µg per mL, of clonidine in the sample obtained from the standard curve;
V is the volume, in mL, of the
Medium; T is the time, in hours; and
A is the area, in cm
2, of the transdermal system.
Tolerances
The release rate of C
9H
9Cl
2N
3 from the Transdermal System, expressed as µg per hour per cm
2 at the times specified, conforms to
Acceptance Table 4.
Time (hours) |
Time for sampling (hours) |
Release rate (µg/h/cm2) |
08 |
8 |
between 7.5 and 16.0 |
824 |
24 |
between 1.5 and 4.6 |
2496 |
96 |
between 1.5 and 4.6 |
96168 |
168 |
between 1.5 and 3.3 |
Drug release 724
Medium:
0.001 M phosphoric acid; 80 mL for systems containing 5 mg or less of clonidine; 200 mL for systems containing more than 5 mg of clonidine.
Times:
8, 24, 96, and 168 hours.
Apparatus 7
Proceed as directed in the chapter, using the transdermal system holder-angled disk (see
Figure 4a). The appropriate size of the holder, 1.42 or 1.98 inches, should be chosen based on the size of the system to prevent overhang. Use 100-mL beakers for
Medium volumes of 80-mL and 300-mL beakers for
Medium volumes of 200 mL. Gently press the transdermal system to a dry, smooth, square piece of cellulose membrane
* , or equivalent, with the adhesive side against the membrane. Attach the membrane/system to a suitable inert sample holder with a Viton O-ring, or equivalent, such that the backing of the system is adjacent to, and centered on, the bottom of the sample holder. Trim the excess of cellulose membrane with scissors. Suspend each sample holder from the arm of a reciprocating shaker such that each system is continuously immersed in a beaker containing the specified volume of
Medium. The filled beakers are weighed and pre-equilibrated to 32.0 ± 0.3
prior to immersing the test sample. Agitate the sample in an up-down motion at a frequency of 30 cycles per minute with an amplitude of 2.0 ± 0.1 cm. The
Medium must be added daily to the beakers during each interval to maintain sample immersion. At the end of each time interval, transfer the test sample to a fresh beaker containing the appropriate volume of
Medium, weighed and pre-equilibrated to 32.0 ± 0.3
.
Determine the amount of C9H9Cl2N3 released by employing the following method.
Mobile phase
Use a filtered and degassed 0.1% solution of triethylamine in a mixture of water and methanol (70:30), adjust with phosphoric acid to a pH of 6.0 ± 0.2. Make adjustments if necessary (see
System Suitability under
Chromatography 621).
System suitability solution
Prepare a solution of USP Clonidine RS in 0.001 M phosphoric acid having a known concentration of about 10 µg per mL.
Standard solutions
Prepare a minimum of four standard solutions of USP Clonidine RS in 0.001 M phosphoric acid having known concentrations of clonidine similar to those of the Test solutions.
Test solutions
At the end of each release interval, allow the beakers to cool to room temperature and make up for evaporative Medium losses by adding Medium to obtain the original weight. Mix.
Chromatographic system (see Chromatography 621)
The liquid chromatograph is equipped with a 220-nm detector and a 4.6-mm × 15-cm column that contains packing L1. The flow rate is about 1.5 mL per minute. Chromatograph the
System suitability solution, and record the peak responses as directed for
Procedure: the tailing factor is not more than 2.0; the capacity factor is not less than 0.5; the column efficiency is not less than 2000 theoretical plates; and the relative standard deviation for replicate injections is not more than 2.0%.
Procedure
Separately inject equal volumes (about 25 µL) of filtered portions of each
Standard solution and
Test solution into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Construct a standard curve of concentration (µg per mL) of clonidine in the
Standard solutions versus peak area by linear regression analysis. The correlation coefficient is not less than 0.995. Calculate the release rate of clonidine by the formula:
CV/TA,
in which
C is the concentration, in µg per mL, of clonidine in the sample obtained from the standard curve;
V is the volume, in mL, of the
Medium; T is the time, in hours; and
A is the area, in cm
2, of the transdermal system.
Tolerances
The release rate of C
9H
9Cl
2N
3 from the Transdermal System, expressed as µg per hour per cm
2 at the times specified, conforms to
Acceptance Table 1 under
Drug Release 724.
Time (hours) |
Time for sampling (hours) |
Release rate (µg/h/cm2) |
08 |
8 |
between 7.5 and 16.0 |
824 |
24 |
between 1.5 and 4.6 |
2496 |
96 |
between 1.5 and 4.6 |
96168 |
168 |
between 1.5 and 3.3 |
(Official April 1, 2006)
Related compounds
Mobile phase, Diluent, System suitability solution, and Chromatographic system
Proceed as directed in the Assay.
Standard solution
Dissolve an accurately weighed quantity of USP Clonidine Related Compound B RS in tetrahydrofuran, and dilute quantitatively, and stepwise if necessary, to obtain a solution having a known concentration of about 1 mg per mL. Prepare a minimum of four
Standard solutions in
Diluent that bracket the expected clonidine related compound B concentration in the sample. The standard concentrations should be within the range of 0.2 to 10.0 µg per mL.
[NOTEStandard solutions are stable for up to 2 days if stored at 4
.
]
Test solution
Use the Assay preparation, prepared as directed in the Assay.
Procedure
Separately inject equal volumes (about 25 µL) of at least three
Standard solutions that will bracket the expected sample concentration range and the
Test solution into the chromatograph, record the chromatograms, and measure the responses for the clonidine related compound B. Calculate the peak response ratios of the analyte, and plot the results. Determine the linear regression equation of the standards by the mean-square method, and record the linear regression equation and the correlation coefficient; it should be not less than 0.995. Determine the concentration of the clonidine related compound B. Calculate the amount, in µg per cm
2, of the clonidine related compound B in the portion of the Transdermal System taken by the formula:
CV/A,
in which
C is the concentration of clonidine related compound B, in µg per mL, obtained from the linear regression analysis;
V is the volume of the
Test solution in mL; and
A is the area of the sample system in cm
2. Not more than 10.0 µg per cm
2 of clonidine related compound B is found.
Assay
Mobile phase
Dissolve 4 mL of triethylamine in 1.6 L of water, and adjust with phosphoric acid to a pH of 3.0. Add 2.4 L of acetonitrile, stir the solution for 30 minutes, filter, and degas. Make adjustments if necessary (see
System Suitability under
Chromatography 621).
Diluent
Prepare a mixture of tetrahydrofuran and methanol (1:1).
System suitability solution
Dissolve an accurately weighed quantity of USP Clonidine RS and USP Clonidine Related Compound B RS in Diluent to obtain a solution having known concentrations of about 250 µg per mL and 10 µg per mL, respectively.
Standard preparation
Dissolve a suitable amount, accurately weighed, of USP Clonidine RS in tetrahydrofuran to obtain a solution having a known concentration of about 1 mg per mL. Prepare a minimum of four
Standard preparations in
Diluent that bracket the expected clonidine concentration in the sample. The standard concentrations should be within the range of 50 to 300 µg per mL.
[NOTEStandard preparations are stable for up to 2 days if stored at 4
.
]
Assay preparation
Remove each Transdermal System from its package, discard the release liner from each system, and transfer into a 50-mL centrifuge tube with a Teflon-lined screw cap. Add the appropriate volume of tetrahydrofuran as listed in the table below.
For systems containing about 2.5 mg
of clonidine |
7.0 mL |
For systems containing about 5.0 mg
of clonidine |
14.0 mL |
For systems containing about 7.5 mg
of clonidine |
21.0 mL |
Mix vigorously on a vortex mixer until the systems are washed down and fully submerged in the tetrahydrofuran. Let the systems soak in tetrahydrofuran for about 5 minutes, and mix on a vortex mixer until the systems are completely delaminated. Allow the systems to remain submerged for an additional 60 minutes, mixing on a vortex mixer every 30 minutes. Add methanol in a volume equal to the volume of tetrahydrofuran, and mix vigorously on a vortex mixer. The solution turns milky. Centrifuge for 10 minutes at 2000 rpm. Use the supernatant as the
Assay preparation.
Chromatographic system (see Chromatography 621)
The liquid chromatograph is equipped with a detector capable of measuring at 210 nm and 242 nm, and a 4.6-mm × 15-cm column that contains packing L10. The flow rate is about 1 mL per minute. The detector is programmed initially to 242 nm and 210 nm after the elution of the clonidine peak but prior to the elution of the clonidine related compound B. The relative retention time is 1.0 for clonidine and 1.5 for clonidine related compound B. Chromatograph the
System suitability solution, and record the peak responses as directed for
Procedure: the resolution,
R, between clonidine and clonidine related compound B is not less than 2.0; the capacity factor,
k ¢, is not less than 0.6 for clonidine; the tailing factor for both clonidine and clonidine related compound B is not more than 3.0; and the relative standard deviation of the clonidine peak area for replicate injections is not more than 2.0%.
Procedure
Separately inject equal volumes (about 25 µL) of at least three
Standard solutions that will bracket the expected sample concentration range and the
Assay preparation into the chromatograph, record the chromatograms, and measure the responses for the clonidine. Calculate the peak response ratios of the analyte, and plot the results. Determine the linear regression equation of the standards by the mean-square method, and record the linear regression equation and the correlation coefficient; it should be not less than 0.995. Calculate the amount, in mg, of C
9H
9Cl
2N
3 in the Transdermal System taken by the formula:
CV/1000,
in which
C is the concentration of clonidine, in µg per mL, obtained from the linear regression analysis; and
V is the volume of the
Assay preparation in mL per sample system.