Procedure
Inject 1 µL of the
Assay preparation and the
Standard preparation, successively, into a suitable gas chromatograph fitted with a flame-ionization detector. Under typical conditions, the instrument contains a 3-mm × 1.2-m glass column packed with 1% (w/w) phase G6 on packing S1AB. The column temperature is maintained at 260
and the helium carrier gas flows at 50 mL per minute. In a suitable chromatogram, the resolution factor,
R (see
Chromatography 621), is not less than 3 between the internal standard and testosterone cypionate peaks, and five replicate injections of a single
Standard preparation show a coefficient of variation of not more than 2% in the peak area ratio of testosterone cypionate to internal standard. Measure the areas under the peaks for testosterone cypionate and cholesteryl caprylate in each chromatogram. Calculate the ratio,
RU, of the area of the testosterone cypionate peak to the area of the internal standard peak in the chromatogram from the
Assay preparation, and similarly calculate the ratio,
RS, in the chromatogram from the
Standard preparation. Calculate the quantity, in mg, of C
27H
40O
3 in the portion of Testosterone Cypionate taken by the formula:
W(RU / RS),
in which
W is the weight, in mg, of
USP Testosterone Cypionate RS in the
Standard preparation, and the other terms are as defined therein.