Procedure
(see
Chromatography 621)Inject equal volumes (about 50 µL) of the
Standard preparation and the
Assay preparation into a chromatograph equipped with a 235-nm detector and a 3.9-mm × 30-cm column that contains packing L1. Calculate the quantity, in mg, of the
Sum of quinidine gluconate and dihydroquinidine gluconate in each mL of the Injection taken by the formula:
(5000 / 3)(C / V)(rb,U + rd,U) / (rb,S + rd,S),
in which
C is the concentration, in mg per mL, of
USP Quinidine Gluconate RS in the
Standard preparation,
V is the volume, in mL, of Injection taken,
rb,U and
rb,S are the peak responses of quinidine obtained from the
Assay preparation and the
Standard preparation, respectively, and
rd,U and
rd,S are the peak responses of dihydroquinidine obtained from the
Assay preparation and the
Standard preparation, respectively.