Chromatographic purity
Test solution
Proceed as directed for Assay preparation in the Assay.
Procedure
Inject a volume (about 1 µL) of 
Test solution into the chromatograph, record the chromatogram, and measure the peak responses. Calculate the percentage of each impurity in the portion of Avobenzone taken by the formula: 
100(rI / rS),
in which 
rI is the response of each individual peak, other than the avobenzone peak, in the chromatogram of the 
Test solution; and 
rS is the sum of the responses of all of the peaks in the chromatogram of the 
Test solution: not more than 3.0% of any individual impurity is found, and the sum of all of the impurities is not more than 4.5%.
 
 
Assay
Standard preparation
Dilute an accurately measured quantity of 
USP Avobenzone RS in acetone, and dilute quantitatively, and stepwise if necessary, with acetone to obtain a solution having a known concentration of about 50 mg per mL.
 
Assay preparation
Transfer about 500 mg of Avobenzone, accurately weighed, to a 10-mL volumetric flask, dilute with acetone to volume, and mix.
Chromatographic system (see Chromatography 
621
)
The gas chromatograph is equipped with a flame-ionization detector and a 0.32-mm × 25-m fused silica capillary column coated with phase G1. The column temperature is maintained at about 200

 until the time of injection, then increased at a rate of 4

 per minute to 280

. The injection port temperature is maintained at 200

, and the detector temperature is maintained at about 280

. Helium is used as the carrier gas. Chromatograph the 
Standard preparation, and record the peak responses as directed for 
Procedure: the resolution, 
R, between avobenzone and any adjacent peak is not less than 1.0; and the relative standard deviation for replicate injections is not more than 2.0%.
 
Procedure
Separately inject equal volumes (about 1 µL) of the 
Standard preparation and the 
Assay preparation into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Calculate the quantity, in mg, of C
20H
22O
3 in the portion of Avobenzone taken by the formula: 
10C(rU / rS),
in which 
C is the concentration, in mg per mL, of 
USP Avobenzone RS in the 
Standard preparation; and 
rU and 
rS are the peak responses obtained from the 
Assay preparation and the 
Standard preparation, respectively.