Identification
B:Ultraviolet Absorption 197U
Spectral range:
450 to 200 nm.
Solution
To a 10-mL volumetric flask containing 14 mg of Nifedipine add 1.0 mL of chloroform, dilute with methanol to volume, and mix. Pipet a 1.0-mL aliquot of the solution into a 100-mL volumetric flask, dilute with methanol to volume, and mix.
Medium:
methanol.
C:
The retention time of the major peak in the chromatogram of the Assay preparation corresponds to that in the chromatogram of the Standard preparation, as obtained in the Assay.
Chloride and Sulfate
To 5.0 g of Nifedipine in a 140-mL beaker add 4.0 mL of 6 N acetic acid and 46 mL of water. Bring carefully to a boil on a hot plate, cool, and filter through paper free of chloride and sulfate. Use this Nifedipine filtrate for the following tests.
Chloride
Pipet 2.5 mL of the Nifedipine filtrate into a 50-mL color-comparison tube, and add 12.5 mL of water. Into a matched color-comparison tube pipet 10 mL of a Standard solution containing 8.2 µg of sodium chloride per mL corresponding to 5 µg of chloride per mL, add 5.0 mL of water, and mix. To each tube add 0.15 mL of 0.3 M nitric acid and 0.3 mL of
silver nitrate TS, and mix: the opalescence exhibited by the Nifedipine filtrate does not exceed that of the Standard solution (0.02%).
Sulfate
Pipet into each of two 50-mL matched color-comparison tubes 1.5 mL of sulfate solution consisting of sufficient potassium sulfate dissolved in water to obtain a sulfate concentration of 10 µg per mL. To each tube add, successively and with continuous shaking, 0.75 mL of alcohol, 0.5 mL of a 6.1% aqueous solution of barium chloride, and 0.25 mL of 6 N acetic acid. Shake for an additional 30 seconds. Pipet into one tube, designated the Standard tube, 15 mL of the sulfate solution. Pipet into the other tube, designated the Specimen tube, 3 mL of the Nifedipine filtrate and 12 mL of water: the turbidity exhibited by the Specimen tube does not exceed that of the Standard tube (0.05%).
Related compounds
[NOTEProtect the
Standard nifedipine solution and the
Test preparation from actinic light. Conduct this test promptly after preparation of the
Standard nifedipine solution and the
Test solution.
]
Mobile phase
Prepare as directed in the Assay.
Standard nifedipine solution
Dissolve an accurately weighed quantity of
USP Nifedipine RS in methanol (about 1 mg per mL), and dilute quantitatively with
Mobile phase to obtain a solution having a known concentration of about 0.3 mg per mL.
Reference solution 1
Dissolve an accurately weighed quantity of
USP Nifedipine Nitrophenylpyridine Analog RS in methanol (about 1 mg per mL), and dilute quantitatively with
Mobile phase to obtain a solution having a known concentration of about 0.6 µg per mL.
Reference solution 2
Dissolve an accurately weighed quantity of
USP Nifedipine Nitrosophenylpyridine Analog RS in methanol (about 1 mg per mL), and dilute quantitatively with
Mobile phase to obtain a solution having a known concentration of about 0.6 µg per mL.
Standard solution
Transfer 5.0 mL of each of the two Reference solutions to a container, add 5.0 mL of Mobile phase, and mix.
Test solution
Prepare as directed for the Assay preparation in the Assay.
System suitability solution
Mix equal volumes of the Standard nifedipine solution and of each of the two Reference solutions.
Chromatographic system
Prepare as directed in the Assay. Chromatograph the System suitability preparation, and record the peak responses as directed for Procedure: the resolution, R, between the nitrophenylpyridine analog and nitrosophenylpyridine analog peaks is not less than 1.5; the resolution, R, between the nitrosophenylpyridine analog and nifedipine peaks is not less than 1.0; and the relative standard deviation of the response for each analog in replicate injections is not more than 10%. The relative retention times are about 0.8 for the nitrophenylpyridine analog, about 0.9 for the nitrosophenylpyridine analog, and 1.0 for nifedipine.
Procedure
Separately inject equal volumes (about 25 µL) of the
Standard solution and the
Test solution into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Calculate the quantity, in mg, of each related compound in the portion of Nifedipine taken by the formula:
250C(rU / rS),
in which
C is the concentration, in mg per mL, of the appropriate USP Nifedipine Analog RS, in the
Standard solution; and
rU and
rS are the peak responses for the corresponding related compound obtained from the
Test solution and the
Standard solution, respectively. Not more than 0.2% of each of dimethyl 4-(2-nitrophenyl)-2,6-dimethylpyridine-3,5-dicarboxylate and dimethyl- 4-(2-nitrosophenyl)-2,6-dimethylpyridine-3,5-dicarboxylate, corresponding to Nifedipine Nitrophenylpyridine Analog and Nifedipine Nitrosophenylpyridine Analog, respectively, is found.
Assay
[NOTEProtect the
Standard preparation and the
Assay preparation from actinic light. Conduct the
Assay promptly after preparation of the
Standard preparation and the
Assay preparation.
]
Mobile phase
Prepare a suitable mixture of water, acetonitrile, and methanol (50:25:25), and degas. Make adjustments if necessary (see
System Suitability under
Chromatography 621).
Standard preparation
Dissolve an accurately weighed quantity of
USP Nifedipine RS in methanol (about 1 mg per mL), and quantitatively dilute with
Mobile phase to obtain a solution having a known concentration of about 0.1 mg per mL.
Assay preparation
Transfer about 25 mg of Nifedipine, accurately weighed, to a 250-mL volumetric flask. Dissolve in 25 mL of methanol, dilute with Mobile phase to volume, and mix to obtain a solution having a concentration of about 0.1 mg per mL.
Chromatographic system (see Chromatography 621)
The liquid chromatograph is equipped with a 235-nm detector and a 4.6-mm × 25-cm column that contains 5-µm packing L1. The flow rate is about 1.0 mL per minute. Chromatograph the
Standard preparation, and record the peak responses as directed for
Procedure: the column efficiency is not less than 4000 theoretical plates; the tailing factor is not more than 1.5; and the relative standard deviation for replicate injections is not more than 1.0%.
Procedure
Separately inject equal volumes (about 25 µL) of the
Standard preparation and the
Assay preparation into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Calculate the quantity, in mg, of C
17H
18N
2O
6 in the portion of Nifedipine taken by the formula:
250C(rU / rS),
in which
C is the concentration, in mg per mL, of
USP Nifedipine RS in the
Standard preparation; and
rU and
rS are the peak responses obtained from the
Assay preparation and the
Standard preparation, respectively.