Assay
Standard preparation
Dissolve an accurately weighed quantity of
USP Homosalate RS in chloroform, and dilute quantitatively, and stepwise if necessary, with chloroform to obtain a solution having a known concentration of about 5.0 mg per mL.
Assay preparation
Dissolve about 50 mg of Homosalate, accurately weighed, in 10 mL of chloroform, and mix.
Chromatographic system (see Chromatography 621)
The gas chromatograph is equipped with a flame-ionization detector and a 0.53-mm × 30-m column coated with a 1-µm film of G27. Hydrogen is used as the carrier gas at a flow rate of about 6 mL per minute. The chromatograph is programmed as follows. Initially the temperature of the column is equilibrated at 70
; upon injection the temperature is increased at a rate of 6
per minute to 220
, and is held at 220
for 5 minutes. The injection port temperature is maintained at 270
, and the detector is maintained at 270
. Chromatograph the
Standard preparation, and record the peak responses as directed for
Procedure: the relative retention times are about 0.98 for the
cis-isomer and 1.0 for the
trans-isomer; the resolution,
R, between the
cis-isomer and the
trans-isomer is not less than 1.0; and the relative standard deviation for replicate injections is not more than 2.0%.
Procedure
Separately inject equal volumes (about 1 µL) of the
Standard preparation and the
Assay preparation into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Calculate the quantity, in mg, of C
16H
22O
3 in the portion of Homosalate taken by the formula:
10C(rU / rS),
in which
C is the concentration, in mg per mL, of
USP Homosalate RS in the
Standard preparation; and
rU and
rS are the sums of the peak responses for the
cis- and
trans-isomers obtained from the
Assay preparation and the
Standard preparation, respectively.