Chromatographic purity
Internal standard solution
Transfer about 500 mg of adamantane, accurately weighed, to a 10-mL volumetric flask, dissolve in and dilute with dichloromethane to volume, and mix.
Test solution
Transfer about 1.0 g of Amantadine Hydrochloride, accurately weighed, to a separator. Add 20 mL of 5.0 N sodium hydroxide and 18 mL of dichloromethane, and shake for 10 minutes. Remove the water layer, dry the organic layer by swirling with anhydrous sodium sulfate, and allow to stand for a few minutes to ensure that all remaining water has been removed. Filter, collect the filtrate in a 20-mL volumetric flask, add 0.2 mL of Internal standard solution, and dilute with dichloromethane to volume.
Standard solution
Transfer about 10 mg of
USP Amantadine Hydrochloride RS, accurately weighed, to a separator. Add 20 mL of 5.0 N sodium hydroxide and 18 mL of dichloromethane, and shake for 10 minutes. Remove the water layer, dry the organic layer by swirling with anhydrous sodium sulfate, and allow to stand for a few minutes to ensure that all remaining water has been removed. Filter, collect the filtrate in a 20-mL volumetric flask, add 0.2 mL of
Internal standard solution, and dilute with dichloromethane to volume.
Chromatographic system (see Chromatography 621)
The gas chromatograph is equipped with a flame-ionization detector and a 0.53-mm × 30-m fused-silica column coated with 1.0-µm G27 stationary phase. The carrier gas is helium, flowing at a rate of about 4 mL per minute, and the split flow is about 200 mL per minute with a split flow ratio of 50:1. Initially the temperature of the column is equilibrated at 70
for 5 minutes, then the temperature is increased linearly at a rate of 10
per minute to 250
, and maintained at 250
for at least 17 minutes. The injection port temperature is maintained at 220
, and the detector temperature is maintained at 300
. Chromatograph the
Standard solution, and record the peak responses as directed for
Procedure: the relative retention times are about 0.7 for adamantane and 1.0 for amantadine hydrochloride; the resolution,
R, between adamantane and amantadine hydrochloride is not less than 20, and the relative standard deviation for replicate injections determined from the peak response ratios of amantadine to adamantane is not more than 5.0%.
Procedure
Separately inject equal volumes (about 2 µL) of the
Standard solution and the
Test solution into the chromatograph, record the chromatograms, and measure all of the peak responses. Calculate the percentage of each impurity in the portion of Amantadine Hydrochloride taken by the formula:
100(Ri / RS)(WS / WU),
in which
Ri is the peak response ratio of each impurity to adamantane obtained from the
Test solution; RS is the peak response ratio of amantadine to adamantane obtained from the
Standard solution; WS is the weight, in mg, of
USP Amantadine Hydrochloride RS taken to prepare the
Standard solution; and
WU is the weight, in mg, of Amantadine Hydrochloride taken to prepare the
Test solution: not more than 0.3% of any individual impurity is found; and not more than 1.0% of total impurities is found.
Assay
Dissolve about 120 mg of Amantadine Hydrochloride, accurately weighed, in a mixture of 30 mL of glacial acetic acid and 10 mL of
mercuric acetate TS, and titrate with 0.1 N perchloric acid VS, determining the endpoint potentiometrically, using suitable electrodes. Perform a blank determination, and make any necessary correction. Each mL of 0.1 N perchloric acid is equivalent to 18.77 mg of C
10H
17N·HCl.